30ECA64 Large Bone Orthopedic Surgical Motor

We respect your privacy. Please see our Privacy Notice for more information.


By contacting us you accept that your details will be used by Portescap to respond to your inquiry.
We respect your privacy. Please see our Privacy Notice for more information.

Large Bone Orthopedic Surgical Motor (30ECA64)

30ECA64 Large Bone Orthopedic Surgical Motor

30ECA64 Large Bone Orthopedic Surgical Motor

30ECA64 Autoclavable Slotless Brushless DC Motor

The Ultra 30ECA64 brushless DC slotless motor is optimized for large bone drill and saw surgical applications. Featuring a 4.7mm cannulated shaft, motor torque and speed make this an ideal solution for powered surgical hand tools that are used for joint replacement surgeries of the knee, hip and shoulder.

Product Highlights

  • Able to withstand 500+ autoclave cycles
  • Optimized for use in surgical powered devices
  • Cannulated shaft version for surgical power drills
  • Non-cannulated shaft version for surgical saws
  • Lightweight and low vibration for increased surgeon tactile feel

Customization Available

  • Coil modifications - resistance and inductance
  • Connectors
  • Custom shafts - flats, knurling, diameter, length
  • Mechanical integration upon request (housing, front & rear flange, shaft)
  • Without hall sensors
Part Number 30ECA64 8B 5 30ECA64 8B 6 30ECA64 8B 5 cannulated 30ECA64 8B 6 cannulated
Motor Diameter (in) 1.18 1.18 1.18 1.18
Motor Diameter (mm) 30 30 30 30
Ambient Operating Temperature Range (°C) -30 to +135 -30 to +135 -30 to +135 -30 to +135
Ambient Operating Temperature Range (°F) -22 to +275 -22 to +275 -22 to +275 -22 to +275
Ambient Storage Temperature Range (°C) -40 to +135 -40 to +135 -40 to +135 -40 to +135
Ambient Storage Temperature Range (°F) -40 to +275 -40 to +275 -40 to +275 -40 to +275
Axial Static Force w/o Shaft Support, Max (lbs) 30 30 30 30.1243926
Axial Static Force w/o Shaft Support, Max (N) 134 134 134 134
Back EMF Constant (V/1000 rpm) 0.69 0.9 0.69 0.9
Ball Bearings Preload (N) 13.2 13.2 13.2 13.2
Continuous Current, Max (A) 18.5 15.6 18.5 15.6
Continuous Mechanical Power, Max (@25°C) (W) 165 142 165 142
Continuous Torque, Max (mNm) 121.4 133.5 121.4 133.5
Continuous Torque, Max (oz-in) 17.2 18.9 17.12 18.9
Electrical Time Constant (ms) 0.23 0.24 0.23 0.24
Housing Material Stainless Steel Stainless Steel Stainless Steel Stainless Steel
Hall Sensor Electrical Phasing (electrical °) 120 120 120 120
Inductance - Phase to Phase (mH) 0.0077 0.011 0.0077 0.011
Internal Resistance - Phase to Phase (Ohm) 0.0336 0.045 0.0336 0.045
Line to Line Resistance (Ohm) 0.0426 0.0548 0.0426 0.0548
Mechanical Time Constant (ms) 2.1 1.6 2.1 1.6
Motor Constant (mNm/W½) 35.9 40.4 35.9 40.4
Motor Constant (oz-in/W½) 5.08 5.721 5.083 5.72
Motor Regulation (10³/Nms) 0.777 0.612 0.777 0.612
Motor Speed, Max (rpm) 30000 30000 30000 30000
No-Load Current, Typical (mA) 700 700 700 700
No-Load Speed (rpm) 16100 16488 16100 16488
Nominal Voltage (V) 11.1 14.8 11.1 14.8
Optimization Direction Symmetrical Symmetrical Symmetrical Symmetrical
Rotor Inertia (gcm²) 26.6 26.6 26.6 26.6
Rotor Inertia (oz-in-sec² 10ˉ⁶) 376 377 377 377
Thermal Resistance (Slotless) (°C/W) 7.9 7.9 7.9 7.9
Thermal Time Constant (s) 1310 1310 1310 1310
Torque Constant (mNm/A) 6.6 8.6 6.6 8.6
Torque Constant (oz-in/A) 0.935 1.218 0.935 1.22
Weight (g) 256 256 256 256
Weight (oz) 9 9.030144 9 9
Winding Temperature, Max (°C) 155 155 155 155
Winding Temperature, Max (°F) 311 311 311 311
Autoclave Cycles 500 500 500 500
Continuous Stall Torque (mNm) 121.4 133.5 121.4 133.5
Continuous Stall Torque (oz-in) 17.19 18.91 17.19 18.91
Ratio - - - -
Length (in) 2.52 2.52 2.52 2.512
Length (mm) 64 64 64 64
Feedback Type Sensored Sensored Sensored Sensored
Number of Poles 4 pole 4 pole 4 pole 4 pole